Comparison of four methods of surface roughness assessment of corneal stromal bed after lamellar cutting - Ecole Nationale des Travaux Publics de l'Etat
Article Dans Une Revue Biomedical optics express Année : 2017

Comparison of four methods of surface roughness assessment of corneal stromal bed after lamellar cutting

Résumé

Corneal lamellar cutting with a blade or femtosecond laser (FSL) is commonly used during refractive surgery and corneal grafts. Surface roughness of the cutting plane influences postoperative visual acuity but is difficult to assess reliably. For the first time, we compared chromatic confocal microscopy (CCM) with scanning electron microscopy, atomic force microscopy (AFM) and focus-variation microscopy (FVM) to characterize surfaces of variable roughness after FSL cutting. The small area allowed by AFM hinders conclusive roughness analysis, especially with irregular cuts. FVM does not always differentiate between smooth and rough surfaces. Finally, CCM allows analysis of large surfaces and differentiates between surface states.
Fichier principal
Vignette du fichier
boe-8-11-4974.pdf (8.85 Mo) Télécharger le fichier
Origine Fichiers éditeurs autorisés sur une archive ouverte

Dates et versions

hal-01634361 , version 1 (28-10-2024)

Identifiants

Citer

Clotilde Jumelle, Alina Hamri, Gregory Egaud, Cyril Mauclair, Stéphanie Reynaud, et al.. Comparison of four methods of surface roughness assessment of corneal stromal bed after lamellar cutting. Biomedical optics express, 2017, 8 (11), pp.4974-4986. ⟨10.1364/BOE.8.004974⟩. ⟨hal-01634361⟩
47 Consultations
0 Téléchargements

Altmetric

Partager

More